• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OGDEN MANUFACTURING PLANT OPTIFLUX 180NRE DIALYZER FINISHED ASSY. DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OGDEN MANUFACTURING PLANT OPTIFLUX 180NRE DIALYZER FINISHED ASSY. DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 0500318E
Device Problem Fluid Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2018
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.  .
 
Event Description
A user facility clinical manager reported that a dialyzer blood leak occurred immediately after the initiation of the patient¿s hemodialysis (hd) treatment. The blood leak was noted as being an internal blood leak. The leak was not visually observed. The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm. Blood leak test strips were used and tested positive for the presence of blood. The patient¿s estimated blood loss (ebl) was approximately 250ml. There was no patient injury, adverse events, or medical intervention required as a result of this event. The patient was restarted on a new machine and treatment completed successfully with new supplies. The complaint device was reported to have been discarded and is not available to be returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameOPTIFLUX 180NRE DIALYZER FINISHED ASSY.
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
475 west 13th street
ogden UT 84404
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key8161864
MDR Text Key130330802
Report Number1713747-2018-00488
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K002761
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/14/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/30/2021
Device Catalogue Number0500318E
Device Lot Number18LU07010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Device Age MO
Event Location No Information
Date Manufacturer Received01/11/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/15/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-