• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES GORE VIABAHN® ENDOPROSTHESIS - 3; NIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES GORE VIABAHN® ENDOPROSTHESIS - 3; NIP Back to Search Results
Catalog Number VBH100502A
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 12/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the manufacturing records for the device verified the lot met all pre-release specifications.
 
Event or Problem Description
The following information was reported to gore: on (b)(6) 2018 a patient was undergoing a procedure to reline a previously implanted stent in an av access application in the subclavian vein with a gore® viabahn® endoprosthesis.The device was attempted to be advanced within the previously implanted stent, but it was not possible to do so.The undeployed device was then withdrawn and as it was being withdrawn it became stuck on the edge of the sheath, causing it to accordion on the delivery catheter.The sheath and device were then removed.As they were removed, some vessel damage occurred which required a cut-down and some additional suturing.A new 10fr sheath was then inserted and the case was completed with a non-gore stent.The patient did well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE VIABAHN® ENDOPROSTHESIS - 3
Common Device Name
NIP
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key8162024
Report Number2017233-2018-00763
Device Sequence Number19364752
Product Code PFV
UDI-Device Identifier00733132623211
UDI-Public00733132623211
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P040037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/25/2020
Device Catalogue NumberVBH100502A
Device Lot Number17578144
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
-
-