BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3853
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/15/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a shaft break occurred.The 90% stenosed target lesion was located in the moderately tortuous and moderately calcified mid right coronary artery (rca).A 6mm x 2.50mm wolverine coronary cutting balloon was being advanced in to the guide catheter when the shaft broke approximately 20 cm from the hub.The wolverine catheter never entered the patient.The procedure was completed with a different device.No patient complications were reported in relation to this event.
 
Additional Manufacturer Narrative
Device evaluated by mfr: based on hazards identified, the following attributes were considered during analysis: the wolverine device was received and analyzed.A complete break was identified in the hypotube the device.A visual and microscopic examination was performed on the returned device.A visual and tactile examination identified a complete break in the hypotube of the device.The break was located at approximately 415mm distal to the strain relief.The hypotube was also noted to be kinked at several locations along its length.This type of damage is consistent with excessive force being applied to the delivery system.No issues were noted with the hypotube that may have contributed to the damage identified.A microscopic examination identified no damage to the tip of the device.A visual and microscopic examination identified no damage to the markerbands or blades of the device.All blades were present and fully bonded to the balloon material.A visual examination identified that the balloon had not been subjected to positive pressure.No issues were noted with the balloon that may have potentially contributed to the complaint incident.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that a shaft break occurred.The 90% stenosed target lesion was located in the moderately tortuous and moderately calcified mid right coronary artery (rca).A 6mm x 2.50mm wolverine coronary cutting balloon was being advanced in to the guide catheter when the shaft broke approximately 20 cm from the hub.The wolverine catheter never entered the patient.The procedure was completed with a different device.No patient complications were reported in relation to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON
Common Device Name
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8162039
Report Number2134265-2018-63861
Device Sequence Number19363144
Product Code NWX
UDI-Device Identifier08714729888154
UDI-Public08714729888154
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P950020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/30/2020
Device Model Number3853
Device Catalogue Number3853
Device Lot Number0022191951
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2018
Initial Date Received by Manufacturer 11/16/2018
Supplement Date Received by Manufacturer12/20/2018
Initial Report FDA Received Date12/13/2018
Supplement Report FDA Received Date01/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-