• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SEE H.10 DELTAXSFT10 2MM X 6CM; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SEE H.10 DELTAXSFT10 2MM X 6CM; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number DLX100206
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Procode is krd/hcg.(b)(4).(b)(6).[conclusion]: the healthcare professional reported that during the coil embolization procedure to treat a ruptured cerebral aneurysm at the anterior communicating artery (acom), the 2mm x 6cm deltaxsft 10 coil (dlx100206 / s13960) was removed from its packaging and the physician attempted to insert it into the stryker sl-10® microcatheter, but the coil was impeded in the microcatheter.The coil was replaced with a second 2mm x 6cm deltaxsft 10 coil from the same lot and the same issue occurred.This second coil was replaced, and the procedure continued to completion without further issues or delay.There was no report of patient injury or complication; the complaint product was new and was stored per labeling instructions.The procedure was conducted in accordance with the instruction for use (ifu) with constant flush maintained at all time.There was no visible defect or damage noted on the product prior to the event.There was no unintended detachment observed in the vessel or in the microcatheter.The 2mm x 6cm deltaxsft 10 coil was returned for evaluation and analysis.The investigational finding is documented below.Investigation summary: the 2mm x 6cm deltaxsft 10 coil was returned in the same package as the second coil from the same lot.There was no identification separating which was the first coil and which was the second coil documented in the complaint.The devices were marked as unit ¿a¿ and ¿b¿ upon analysis in the product analysis lab.Unit ¿b¿ was assigned to this investigation.The distal end of the embolic coil was seen in the green introducer.Blood was observed throughout the green and translucent sections of the introducer.Kinks were observed on the device positioning unit (dpu) core wire at approximately 27 cm and 90 cm from the proximal end of the device.The device was then inspected under a microscope.The embolic coil distal ball tip was present and intact.The articulating joint was seen, and it appeared that the embolic coil had been pressed into the distal outer sheath.Blood was seen obscuring the view of the embolic coil and distal outer sheath, and as a result, the shape of the distal outer sheath could not be determined.A kink was observed on the embolic coil.The v-notch of the resheathing tool was seen undamaged.An attempt was made to advance the device, but the dpu core wire and embolic coil could not move in the introducer due to the resistance caused by blood.The device was unable to be advanced through a microcatheter due to the condition in which it was returned.During the review of the lot history records for lot s13960, it was noted that 4 units were rejected due to damaged coils which is a condition that may be associated with the type of failure mode as reported in the complaint.However; the device history record (dhr) review confirmed that all product rejected during manufacturing were identified as scrap and properly accounted for.No other issues were noted that were considered potentially related to the reported complaint.Investigation conclusion: the reported issue that the coil delivery system (dcs) was impeded in the microcatheter was not confirmed.The returned device could not be tested due to the condition in which it was returned.Dried blood impeded the movement of the dpu core wire and the embolic coil through the coil introducer.The stryker sl-10® microcatheter that was used during the procedure was not returned; therefore, a complete analysis could not be performed.The kink observed on the embolic coil is likely caused by the application of force and the resistance caused by the blood that was present.The blood in the coil introducer and surrounding the embolic coil and articulating joint suggests that there was sufficient flush was not maintained.The instruction for use (ifu) states that continuous infusion of an appropriate flush solution is required for optimal performance, and also indicates that the flush should be verified in the event of resistance.Insufficient flush allows blood to back-flow into the microcatheter, which can result in resistance.Devices undergo 100% in-process inspection.The kink observed on the embolic coil would have been caught during inspection at final assembly.It is unlikely that the 2mm x 6cm deltaxsft 10 coil left the manufacturing facility with the damages observed on the embolic coil.Coil damage such as kinking is a known potential issue associated with the use of this device.The ifu contains instructions on proper handling of the device.The reported issues captured in the complaint that the coil delivery system (dcs) was impeded in the microcatheter was not confirmed through evaluation of the returned device due to the condition in which it was returned.Based on the device history record review, it was confirmed that all product rejected during manufacturing were identified as scrap and properly accounted for.The coil was observed to be kinked, which is a reportable malfunction, was observed.The exact cause of the coil kink was not conclusively determined, but the likely cause of the kink observed on the embolic coil is applied force.In addition, it also appears to be related to an insufficient flush that may have resulted in resistance.It is also possible that clinical and procedural factors, including device handling, device manipulation/interaction, aneurysm size/vessel characteristics, device selection, and the concomitant microcatheter, may have contributed to the observed issue during the evaluation of the returned product.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This is one of two (2) with the reported complaint.The associated manufacturer report numbers are 3008114965-2018-00837.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The healthcare professional reported that during the coil embolization procedure to treat a ruptured cerebral aneurysm at the anterior communicating artery (acom), the 2mm x 6cm deltaxsft 10 coil (dlx100206 / s13960) was removed from its packaging and the physician attempted to insert it into the stryker sl-10® microcatheter, but the coil was impeded in the microcatheter.The coil was replaced with a second 2mm x 6cm deltaxsft 10 coil from the same lot and the same issue occurred.This second coil was replaced, and the procedure continued to completion without further issues or delay.There was no report of patient injury or complication; the complaint product was new and was stored per labeling instructions.The procedure was conducted in accordance with the instruction for use (ifu) with constant flush maintained at all time.There was no visible defect or damage noted on the product prior to the event.There was no unintended detachment observed in the vessel or in the microcatheter.The 2mm x 6cm deltaxsft 10 coil was reported as available to be returned for evaluation and analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTAXSFT10 2MM X 6CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
SEE H.10
47709 fremont blvd
fremont CA 94538
MDR Report Key8162192
Report Number3008114965-2018-00836
Device Sequence Number2371144
Product Code KRD
UDI-Device Identifier10886704077336
UDI-Public10886704077336
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Catalogue NumberDLX100206
Device Lot NumberS13960
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/12/2018
Initial Report FDA Received Date12/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-