• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH LATERAL ASSEMBLY, RADIAL STEM IMPLANT, SIZE 4; ELBOW JOINT RADIAL (HEMI-ELBOW) POLYMER PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH LATERAL ASSEMBLY, RADIAL STEM IMPLANT, SIZE 4; ELBOW JOINT RADIAL (HEMI-ELBOW) POLYMER PROSTHESIS Back to Search Results
Catalog Number 3100003
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available it will be provided on a supplemental report.Device was not received.
 
Event or Problem Description
Patient was lifting a box overhead, lateral radial head dislodged from stem.Both head and stem implants were explanted.Competitive product used as replacement.
 
Additional Manufacturer Narrative
The reported event that lateral assembly, radial stem implant, size 4 was alleged of 'component loosened' could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event such as the x-rays of the patient pre- and post- operatively, as well as the patient information and the affected device must be available in order to determine the exact root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
Patient was lifting a box overhead, lateral radial head dislodged from stem.Both head and stem implants were explanted.Competitive product used as replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATERAL ASSEMBLY, RADIAL STEM IMPLANT, SIZE 4
Common Device Name
ELBOW JOINT RADIAL (HEMI-ELBOW) POLYMER PROSTHESIS
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key8162197
Report Number0008031020-2018-01057
Device Sequence Number1997791
Product Code JDB
UDI-Device Identifier00886385022703
UDI-Public00886385022703
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K070236
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date12/31/2019
Device Catalogue Number3100003
Device Lot Number10187-L04
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/15/2018
Supplement Date Received by Manufacturer01/24/2019
Initial Report FDA Received Date12/13/2018
Supplement Report FDA Received Date02/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
-
-