• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMERICAN MEDICAL SYSTEMS, INC. SUREFLEX FIBERS; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMERICAN MEDICAL SYSTEMS, INC. SUREFLEX FIBERS; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number S-LLF200TG
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record (dhr) confirmed that the devices met all material, assembly and performance specifications.Analysis of the device revealed the fiber was returned in two pieces.The distal end of the glass fiber exhibits no signs of usage.The fiber is broken within the fiber blue outer sheath, and detached at 11 feet and 7.5 inches from distal tip.The length of second piece including the connector is 6.5 inches.The fiber was tested with hene laser fixture, aim beam is visible at the break.There is no sign of melting at the break location.An evaluation conclusion code of cause not established was assigned to this investigation.
 
Event or Problem Description
Analysis of the returned device found the fiber broken into two pieces.There were no clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFLEX FIBERS
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC.
3070 orchard drive
san jose CA 95134
MDR Report Key8162336
Report Number2937094-2018-60152
Device Sequence Number11313400
Product Code GEX
UDI-Device Identifier00878953004822
UDI-Public00878953004822
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K140679
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/03/2021
Device Model NumberS-LLF200TG
Device Catalogue NumberS-LLF200TG
Device Lot Number0021917858
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2018
Initial Report FDA Received Date12/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-