| Catalog Number |
314.08 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is synthes sales representative without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2018, a holding sleeve had a missing spring and was unusable.There was no patient or procedure involvement.This report is for one (1) holding sleeve.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A device history record (dhr) review was conducted: part: 314.080, lot: 8818178, manufacturing site: haegendorf, release to warehouse date: 21.Mar.2014.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The customer reported a holding sleeve had missing spring and was unusable.The repair technician reported the coupling sleeve and spring were missing.Missing parts is the reason for repair.The cause of the issue is unknown.The following parts were replaced: coupling sleeve, spring, and holding sleeve.The item was repaired per the inspection sheet, passed synthes final inspection, and will be returned to the customer upon completion of the service and repair process.Finalized service record will be archived.The evaluation was confirmed.The device was deemed serviceable and returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A device history record (dhr) review was conducted: part: 314.080.Lot: 8819178.Manufacturing site: haegendorf.Release to warehouse date: 21.Mar.2014.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|