TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number METRIX AIR
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-18- user has high glucose value test strip udi#: (b)(4).Note: manufacturer contacted customer on (b)(6) 2018, in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
 
Event or Problem Description
Consumer reported complaint for hi display.Friend is calling on behalf of the customer.The expected fasting blood glucose test result range is 150 to 200mg/dl.The customer did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the bedroom.During the call on (b)(6) 2018, a blood test was performed by the customer and produced test results of 286 and 295mg/dl using true metrix air meter.The test strip lot manufacturer's expiration date is 09/16/2019 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history.(b)(6).At the time of the call customer states she was hospitalized in (b)(6) or (b)(6) of 2017.Customer states that went to the hospital due to neck pain, fast pulse, and feet cramped.Customer was told that her sugar was in the 300's and was dehydrated.Customer received iv fluids and dextrose.
 
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) internal report # (b)(4) returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-18- user has high glucose value test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(6) 2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
 
Event or Problem Description
Consumer reported complaint for hi display.Friend is calling on behalf of the customer.The expected fasting blood glucose test result range is 150 to 200mg/dl.The customer did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the bedroom.During the call on 11/20/2018, a blood test was performed by the customer and produced test results of 286 and 295mg/dl using true metrix air meter.The test strip lot manufacturer's expiration date is 09/16/2019 and open vial date is 10/20/18.The meter memory was reviewed for previous test result history.Result 1:305mg/dl, date: (b)(6) 2018, time:12:11pm, fasting; result 2:354mg/dl, date: (b)(6) 2018, time:12:03pm, fasting; result 3:250mg/dl, date: (b)(6) 2018, time:11:54am, fasting; result 4:255mg/dl, date: (b)(6) 2018, time:7:54am, fasting; result 5:254mg/dl, date: (b)(6) 2018, time:10:36am, fasting; result 6:hi, date: (b)(6) 2018, time:5:01pm, non-fasting.At the time of the call customer states she was hospitalized in (b)(6) of 2017.Customer states that went to the hospital due to neck pain, fast pulse, and feet cramped.Customer was told that her sugar was in the 300's and was dehydrated.Customer received iv fluids and dextrose.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8162585
Report Number1000113657-2018-01247
Device Sequence Number1697557
Product Code NBW
UDI-Device Identifier00087701427398
UDI-Public(01)00087701427398
Combination Product (Y/N)N
Initial Reporter StateWA
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 01/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMV2822
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2018
Was the Report Sent to FDA? No
Event Location Home
Initial Date Received by Manufacturer 11/20/2018
Supplement Date Received by Manufacturer11/20/2018
Initial Report FDA Received Date12/13/2018
Supplement Report FDA Received Date01/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY; SECOND THERAPY
-
-