| Model Number |
CQ7568 |
| Medical Device Problem Codes |
Material Frayed (1262); Peeled/Delaminated (1454); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 04/2021).
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure, the pta balloon allegedly ruptured at 8 atm.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.A visual inspection found peeling along the length of the balloon, and frayed fibers on the proximal cone of the balloon.Therefore, the investigation is confirmed for peeling and for frayed fibers.The device was inflated and water was seen exiting the proximal cone.The fibers were stripped away, and a longitudinal rupture was noted in the balloon material.Therefore, the investigation is confirmed for a longitudinal rupture.The definitive root cause for the identified issues could not be determined based upon the available information.An investigation is currently open to address the peeling outer layer issue.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure, the pta balloon allegedly ruptured at 8 atm.There was no reported patient injury.
|
| |
|