• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U3575610
Medical Device Problem Codes Deflation Problem (1149); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Needle Stick/Puncture (2462)
Date of Event 11/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 01/2021).
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly had issues deflating.There was no reported patient injury.
 
Additional Manufacturer Narrative
This supplemental emdr was sent to reflect the changes in the file, due to information received on 1/3/2019.Field b1, adverse event report type, was changed after receipt of new information, thus requiring a supplemental emdr within 30 days.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 01/2021).
 
Event or Problem Description
It was reported that during an angioplasty procedure in the left superficial femoral artery, the pta balloon allegedly had issues deflating.A needle stick was used to puncture and deflation the device.Another device was used to complete the procedure.There was no reported patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the left superficial femoral artery, the pta balloon allegedly had issues deflating.A needle stick was used to puncture and deflation the device.Another device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review:the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.During the visual inspection it was noted that the device was returned in two segments, and the balloon had a tear along its entire length.This damage will be considered incidental as they were not reported and would not have been present at the time of the procedure.Due to the condition of the sample the investigation is inconclusive for the reported deflation issue, as the device was unable to be functionally tested.Based on the evaluation results, the device was unable to be fully functionally tested due to the condition of the sample.Therefore, the definitive root cause for the reported deflation issues could not be determined.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035 PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8162737
Report Number2020394-2018-02204
Device Sequence Number2120082
Product Code LIT
UDI-Device Identifier00801741092985
UDI-Public(01)00801741092985
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU3575610
Device Catalogue NumberU3575610
Device Lot NumberCMCP0315
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2019
Initial Date Received by Manufacturer 11/19/2018
Supplement Date Received by Manufacturer01/03/2019
03/27/2019
Initial Report FDA Received Date12/13/2018
Supplement Report FDA Received Date01/31/2019
04/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-