• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER SEDLINE KIT; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER SEDLINE KIT; MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 26187
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event or Problem Description
The customer reported the psi value was high.No patient impact or consequences were reported.
 
Event or Problem Description
The customer reported the psi value was high.No patient impact or consequences were reported.
 
Additional Manufacturer Narrative
Additional manufacuring narrative: other, other text: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was determined to be fully functional., corrected data: model# corrected from 9513 to 26187.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEDLINE KIT
Common Device Name
MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key8162805
Report Number2031172-2018-00573
Device Sequence Number1606368
Product Code MWI
UDI-Device Identifier00843997009096
UDI-Public00843997009096
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K151644
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26187
Device Catalogue Number9513
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/19/2018
Supplement Date Received by Manufacturer02/01/2019
Initial Report FDA Received Date12/13/2018
Supplement Report FDA Received Date02/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-