| Model Number |
FT18JN40NSE055N |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Occupation: clinical resource manager.
|
| |
|
Event or Problem Description
|
|
Information was received that a smiths medical bivona custom tracheostomy tube cuff leaked while in use with a picu patient.Subsequently, respiratory was called to the bedside to perform an emergent tracheostomy (trach) change out.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One flextend tracheostomy tube with tts cuff was returned for analysis.Visual inspection revealed that the cuff was torn.It was noted to be 2mm across and 9mm in length.The investigation did not reveal a manufacturing defects.Based on the evidence, the complaint was confirmed.However, the root cause is unknown.
|
| |
|