TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problems Mechanical Problem (1384); Obstruction of Flow (2423)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/25/2017
Event Type  malfunction  
Manufacturer Narrative
Device evaluation by manufacturer: a field service engineer (fse) found the filter was possibly clogged and the filter housing was very tight on the filter, restricting flow.The fse checked the fluidic lines and column, installed a new filter, and g8 analyzer operated without any error messages.The fse then proceeded to run quality controls and patient samples, which passed.The g8 analyzer was working in accordance to the manufacturer's expectations; no further action was required.A 13-month complaint and service history review was performed on serial number (b)(4) for similar complaints from 25-aug-2016 through aware date (b)(4) 2017.No other similar complaints were identified during the searched period.G8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.100 pressure high: the pump pressure exceeded the upper limit (15 mpa) set in the pres high parameters.When the filter or column replacement period has been exceeded, first replace the filter or column.If the pressure is still high, remove the inlet and outlet flow line around the column and filter, and determine which part is the cause of the high pressure.Then, contact a technical support representative.If the pressure displayed on the screen is: (a) greater than the pressure on the column inspection report + 4 mpa, then replace the filter.(b) less than the pressure on the column inspection report, then proceed with priming the column.The most probable cause of the reported event was due to a clogged filter.This report is being submitted due to an internal nonconformance identified.(b)(4).
 
Event Description
A customer reported receiving error message "100 pressure high" with their g8 analyzer.The customer reported that pressure was over 21 megapascal (mpa).The technical support specialist (tss) had the customer disconnect the peek tubing on the output side of the pre-filter housing and pressure was 0.1 mpa.The tss then instructed the customer to disconnect the peek tubing on the input side of the column and the pressure was 21 mpa (upper limit is 15 mpa).The customer required service in order to investigate the issue.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting of hemoglobin a1c (hba1c) patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 105-8 623
JA  105-8623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key8162915
MDR Text Key131341220
Report Number8031673-2018-05321
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA12/13/2018
Distributor Facility Aware Date09/25/2017
Device Age7 YR
Event Location Hospital
Date Report to Manufacturer12/03/2018
Date Manufacturer Received09/25/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-