The explanted devices were received for analysis.The lead assembly was returned in one portion with one loose tie down; the electrodes were not returned.Setscrew marks were observed on the connector pin, providing evidence of proper mechanical and electrical contact between the generator and connector pin at one point in time.Abraded openings of both the outer and inner tubing were identified in the returned lead portion.The abraded openings found on the outer silicone tubing, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.With the exception of the abraded openings, the condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.Continuity checks of the returned lead portion were performed, during the visual analysis, and no discontinuities were identified.Note that since the electrode array section was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.These lead malfunctions will be captured in mfg report # 1644487-2019-00414.Analysis of the generator has not been completed to date.No additional, relevant information was received to date.
|