| Model Number |
MMT-723RNAB |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
09/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pump passed the displacement test.Unit received with motor error alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform prime/compromised force sensor system alert test, excessive no delivery test and occlusion test due to motor error alarm.
|
| |
|
Event or Problem Description
|
|
The customer's son was reported via phone call that the insulin pump had motor error.The customer¿s current blood glucose level was 333 mg/dl.The drive support cap was normal.Customer was able to complete insulin pump rewind.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|