| Catalog Number |
209830 |
| Medical Device Problem Codes |
Fracture (1260); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
¿as part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.¿.
|
| |
|
Event or Problem Description
|
|
Impactor handle and impactor platform broke during impaction.Case type: tha.Surgical delay- =15 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1 and final report submitted.Product identification: the reported device was confirmed to be a trident inline offset impactor, p/n 209830, lot 32884, rma 263629.Inspection: visual inspection shows the impactor shaft broken in two separate pieces.Refer to attachment.Product history: review of device history records indicate that 8 devices were accepted into final stock on 07/07/16 with no reported discrepancies.Complaint history review: review of complaint within trackwise database for p/n 209830, lot 32884 shows 0 other complaints related to the failure in this investigation.Nc/capa history review: a review of stryker¿s nc/capa database indicated there have been zero ncs or capas associated with the product and failure mode reported in this event.Conclusion: part inspection confirms reported issue showing a fracture shaft in the distal end of the impactor.
|
| |
|
Event or Problem Description
|
|
Impactor handle and impactor platform broke during impaction.Case type: tha.Surgical delay =15 minutes.
|
| |
|