• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER FORCE TRIAD; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER FORCE TRIAD; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number FORCETRIADZE
Medical Device Problem Code Self-Activation or Keying (1557)
Health Effect - Clinical Code No Information (3190)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the generator was resetting cut and coagulation to one on its own.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The returned sample met specification as received.External inspection found minor scratches and marks on outer casing.Two (2) rear rubber feet were missing and the front two were aged.The reported condition was not confirmed.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORCE TRIAD
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow drive
boulder CO 80301
MDR Report Key8164292
Report Number1717344-2018-01627
Device Sequence Number2026932
Product Code GEI
UDI-Device Identifier10884521133310
UDI-Public10884521133310
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K110268
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFORCETRIADZE
Device Catalogue NumberFORCETRIADZE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2018
Initial Date Received by Manufacturer 11/20/2018
Supplement Date Received by Manufacturer01/17/2019
Initial Report FDA Received Date12/14/2018
Supplement Report FDA Received Date02/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-