Model Number TRUEMETRIX
Medical Device Problem Codes Unable to Obtain Readings (1516); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fatigue (1849); Polydipsia (2604)
Date of Event 11/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not returned for evaluation.Customer declined replacement product at this time.Most likely underlying root cause: mlc-18- user has high glucose value.Test strip udi#: (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Unable to send product letter.No address for customer on file.
 
Event or Problem Description
Consumer reported complaint for e--5 blood glucose test results accompanied by symptoms.The expected fasting blood glucose test result range is undisclosed.The customer did report symptoms of feeling tired and extremely thirsty.Medical attention is not reported as a result of the actual blood glucose results.The product storage location is undisclosed.During the call on (b)(6) 2018, a blood test was performed by the customer and produced test results of e-5 using true metrix meter.The test strip lot manufacturer's expiration date is 09/29/2019, and open vial date is undisclosed.The meter memory was not reviewed for previous test result history.The customer is seeing the e5 after the blood sample has been applied.The e5 error appeared when the customer performed the blood test due to nurse stated that customer does run high and have symptoms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8164628
Report Number1000113657-2018-01249
Device Sequence Number13960007
Product Code NBW
UDI-Device Identifier00021292006051
UDI-Public(01)00021292006051
Combination Product (Y/N)N
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberX56S7
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2018
Initial Report FDA Received Date12/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-