U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA REVERSE SHOULDER SYSTEM GLENOID POLYAXIAL LOCKING SCREW - L30; GLENOID LOCKING SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA REVERSE SHOULDER SYSTEM GLENOID POLYAXIAL LOCKING SCREW - L30; GLENOID LOCKING SCREW Back to Search Results
Catalog Number 04.01.0161
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Preliminary investigation performed by r&d shoulder manager: as reported by the event description, the breakage of the screw may have been caused by a malalignment between the screwdriver and the screw.Batch review performed on 14 december 2018; lot 174744: (b)(4) items manufactured and released on 25 january 2018.Expiration date: 2023-01-15.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
 
Event or Problem Description
During surgery the prongs of the glenoid polyaxial locking screw broke.The prongs and the inner locking screw were removed, the body of the outer locking screw was left in place.The breakage of the screw happened when it came in contact with the metaglene.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE SHOULDER SYSTEM GLENOID POLYAXIAL LOCKING SCREW - L30
Common Device Name
GLENOID LOCKING SCREW
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
MDR Report Key8164768
Report Number3005180920-2018-01007
Device Sequence Number1684378
Product Code PHX
UDI-Device Identifier07630040706483
UDI-Public07630040706483
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K170452
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/15/2023
Device Catalogue Number04.01.0161
Device Lot Number174744
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2018
Initial Report FDA Received Date12/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;