• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 2; CEMENTLESS HIP STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 2; CEMENTLESS HIP STEM Back to Search Results
Catalog Number 01.12.22SN
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 11/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 13 december 2018: lot 134351: (b)(4) items manufactured and released on 08-nov-2013.Expiration date: 2018-09-30.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Clinical evaluation performed by medical affairs director: revision surgery occurred 4, 5 years after primary cementless total hip arthroplasty in a (b)(6) year old woman.Radiographic image provided shows the presence of radiotransparent areas in gruen zones 1, 3,4, 5 and 7 probably due to osteolysis or other type of bone reaction.The reason of this event cannot be determined.
 
Event or Problem Description
The patient came in complaining of pain 4 years and 8 months after the primary.The stem was loose and micro movement had caused heterotopic ossification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 2
Common Device Name
CEMENTLESS HIP STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
MDR Report Key8164773
Report Number3005180920-2018-01001
Device Sequence Number1901202
Product Code JDI
UDI-Device Identifier07630030802621
UDI-Public07630030802621
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K082792
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2018
Device Catalogue Number01.12.22SN
Device Lot Number134351
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2018
Initial Report FDA Received Date12/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-