• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NATUS MANUFACTURING LTD TECA ELITE NEEDLE; TECA ELITE DISPOSABLE CONCENTRIC NEEDLE ELECTRODE,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NATUS MANUFACTURING LTD TECA ELITE NEEDLE; TECA ELITE DISPOSABLE CONCENTRIC NEEDLE ELECTRODE, Back to Search Results
Model Number S53156
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 09/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Customer reported 3 cases where hub separation occurred.It is not known if the 3 cases were of the same part number and lot number, or if the needles were used on the same patient or 3 different patients.This information will be clarified.Justification for not providing below information and applicable sections: part a: patient information - no patient injury initially reported, device malfunction occurred.This will be confirmed.Date of event - date of event unknown, this information will be requested.The medical device is not implantable.Reprocessor name and address - this section is not applicable as the medical device is not a single-use device that was reprocessed or reused on a patient.If remedial action initiated , check type - this section is not applicable as no remedial action was initiated.If action reported to fda under 21 usc 360i (f), list correction / removal reporting number - this section is not applicable as there was no action reported under 21usc 360i(f).
 
Event or Problem Description
Hub separation needle issue.Defective needles, the plastic protector is released when the cable is disconnected and the needle remains connected to the needle holder.Customer has reported 3 cases.
 
Additional Manufacturer Narrative
Customer confirmed no patient injury occurred.The following test was carried out on a qty of 10 needles (retains): hub to cover retention test as per doc-011097 rev q - all needles passed with results within minimum spec of 1.2kgf.Lowest value recorded was 2.610kgf, highest recorded value was 6.9200kgf.Defective needles were returned to middleton however the defective product was not returned to gort from middleton.Email has been sent to the middleton qa manager to make them aware of the situation, due to the defective needles not returning to gort natus was unable to complete a full investigation on the parts.Capa (b)(4) opened to investigate this issue further.
 
Additional Manufacturer Narrative
Customer has reported 3 cases.On initial report it was not known if the 3 cases were of the same part number and lot number, or if the needles were used on the same patient or 3 different patients.This information was clarified by the customer on the (b)(6) 2018.The 3 incidents involved the part number s53156 and lot number 45b/17/t.This occurred after use on 3 different patients.The three incidents will be investigated under complaints: · (b)(4) (ae no.3005581270-2018-00018), · (b)(4) ((ae no.3005581270-2018-00019).Work order (b)(4) was reviewed.No ncr or rework noted.Approved production restart form (b)(4) present in relation to the grinder - full coolant change.Approved run at risk 179 and 180 related to - retest of environmental monitoring carried out in the cleanroom and the refurbishment of the millipore system.All samples taken for assembly in process tests gave acceptable results and were recorded on (b)(4).Readings for hub to cover retention test were within the minimum spec of 1.2kgf ranging between 2.460kgf - 7.020kgf.(b)(4) risk and risk management report - teca disposable needles and value disposable needles and value line dcn was reviewed.There are two potential hazards and hazardous situations associated with physical injury - needle stick, id 25 (needles are sharp, users could pierce themselves) and 31 (cannula & hub assembly separates from outer color cover).The overall risk rating is 3a (severity of hazard - 3- moderate, probability of hazard - a - unlikely).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECA ELITE NEEDLE
Common Device Name
TECA ELITE DISPOSABLE CONCENTRIC NEEDLE ELECTRODE,
Manufacturer (Section D)
NATUS MANUFACTURING LTD
ida business park
gort
galway, connaught H91PD 92
EI  H91PD92
MDR Report Key8165180
Report Number3005581270-2018-00017
Device Sequence Number2791837
Product Code IKT
UDI-Device Identifier00382830017503
UDI-Public00382830017503
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K112034
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/07/2020
Device Model NumberS53156
Device Catalogue NumberS53156
Device Lot Number45B/17/T
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/09/2018
Supplement Date Received by Manufacturer11/09/2018
11/09/2018
Initial Report FDA Received Date12/14/2018
Supplement Report FDA Received Date01/10/2019
02/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-