Model Number 5392
Device Problems Intermittent Capture (1080); Display or Visual Feedback Problem (1184)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/28/2018
Event Type  Injury  
Manufacturer Narrative
Product analysis: analysis was able to confirm the reported error message; the main printed circuit board (pcb) was out of electrical specification.Analysis also noted that two case screws were contaminated.All found defective parts were replaced and all other identified issues were resolved.The device passed final functional and system tests.No other anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) experience a loss of capture and an error code appeared on the screen.The caller requested information outlining the steps required for testing the epg.The caller was provided with the information for testing and was advised to return the epg for service if the testing was out of specification or if intermittent capture was observed.The epg is expected to be returned for service.No patient complications have been reported as a result of this event.It was further reported that the epg was returned for service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key8165236
MDR Text Key130364940
Report Number3004593495-2018-01163
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169725126
UDI-Public00643169725126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/28/2018
Initial Date FDA Received12/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-