The axium pushwire was returned for evaluation without the instant detacher and implant coil.Therefore, any contributing factors from the instant detacher and implant coil could not be assessed.The axium pushwire appeared to be broken on the proximal end.The proximal tip of the pushwire containing the tack weld replacement (twr) crimps was not returned.The pushwire was found to be bent at the proximal end.The pushwire was found to be intact distal to the break indicator at the manual detachment location (via hypotube).Under the microscope, the coin was found to be pulled back from the lumen stop, the shield coil was found to be present, and the implant coil was found to be detached.The outer jacket was then removed to gain access the coin.The coin was measured and was found to be within specifications.The inner diameter of the lumen stop was found to be visually acceptable.The lumen stop inner diameter (id) was measured and found to be within specification.The retainer ring inner diameter (id) was measured and found to be withi n specification.All other subassemblies appeared to be normal and no other anomalies were observed.All other subassemblies appeared to be normal and no other anomalies were observed.Based on the analysis finding, the report of ¿non-detachment¿ could not be confirmed.The axium pushwire was returned for evaluation with the implant coil already detached.The event cause could not be determined.The proximal tip of the pushwire containing the tack weld replacement crimps was not returned.The user is directed to attempt detachment with the first instant detacher 3 times.If the initial instant detacher fails to detach the implant coil, the user is directed to utilize a second instant detacher for 3 attempts.The proximal tip of the pushwire was found to be broken; it is likely due to the attempts to detach the implant coil by the manual method of detachment.It is possible that the pushwire became broken, and proximal tip of the pushwire and the implant coil became lost during shipping back for evaluation.In addition, it is possible that the pushwire became bent during shipping back to medtronic for evaluation as the user did not report the bend in the pushwire at the time of the event.It is likely that the implant coil detached due to the coin position (pulled back from the lumen stop).The review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.All devices are 100% inspected for damages and irregularities during manufacture.Per our instructions for use (ifu): ¿if the coil does not detach after 3 attempts, discard the i.D.(instant detacher) and replace with a new i.D.(instant detacher).In the rare event that the coil does not detach and cannot be removed from the implant delivery pusher, use the following steps for detachment.A.Grip the hypotube approximately 5 cm distal of the positive load indicator at the hypotube break indicator and bend the implant delivery pusher just distal to the hbi 180 degrees.B.Next straighten the pusher back, continue bending and straightening until the pusher tubing opens exposing the release element.C.Gently separate the proximal and distal ends of the open pusher.Then, under fluoroscopy, pull the proximal portion of the implant delivery pusher approximately 2-3 cm to confirm implant detachment per ifu.¿ medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|