Catalog Number 0504990800
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was a breach in the sterile barrier.
 
Additional Manufacturer Narrative
The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Through the investigation it was confirmed that rust was not found on the devices inside the sterile packaging, therefore making this event not reportable.Updated rationale from malfunction to no report required.Manufacture date is not known.Alleged failure: the rusty electrodes can be clearly seen before opening the package.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root causes could be user dissatisfaction.
 
Event or Problem Description
Per investigation results, it is confirmed that rust was not found on the device in the sterile packaging.Therefore making this event not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 22 FR, COLLINGS KNIFE ELECTRODE
Common Device Name
CYSTOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8165772
Report Number0002936485-2018-00719
Device Sequence Number1498067
Product Code FAJ
UDI-Device Identifier37613327055598
UDI-Public37613327055598
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K040390
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number0504990800
Device Lot NumberSTRKN01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2018
Initial Date Received by Manufacturer 11/25/2018
Supplement Date Received by Manufacturer11/25/2018
Initial Report FDA Received Date12/14/2018
Supplement Report FDA Received Date01/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-