Catalog Number PC9004
Health Effect - Clinical Code No Information (3190)
Date of Event 11/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Mfg date: the device manufacture date is not known at this time.However, should it become available, it will be provided in future reports.Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was a thermal event.
 
Additional Manufacturer Narrative
The product was not returned for investigation and the reported failure mode was not confirmed.The reported failure mode will be monitored for future reoccurrence.Alleged failure: customer reported flames between the light lead and scope light post attachment during use.Probable root cause: the reported failure mode is consistent with previous failures in which there was residue left over from sterilization between the laparoscope and light guide connection points.This residue burned up and created smoke, leading to darkening and loss of image.Manufacture date is not known.H3 other text : 81.
 
Event or Problem Description
It was reported that there was a thermal event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINPOINT LIGHT CABLE
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
burnaby
san jose CA 95138
MDR Report Key8165773
Report Number3012345110-2018-00017
Device Sequence Number11127589
Product Code GCJ
UDI-Device Identifier00858701006568
UDI-Public00858701006568
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K150956
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPC9004
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/20/2018
Supplement Date Received by Manufacturer11/20/2018
Initial Report FDA Received Date12/14/2018
Supplement Report FDA Received Date01/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-