| Model Number |
2878-04-101 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the concorde lift driver (287804101) broke off when the surgeon was trying to get additional expansion.The surgeon has used the device several times and is familiar with the torque out design but was trying to get additional expansion after torque out when the tip of the driver shaft broke off in the implant.There was no negative consequence to the patient during the case because of this.Clinical goals were achieved.Patient consequence? no.Is the information being submitted for this complaint all the details that are known/available regarding this event? yes.
|
| |
|