• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 977C165
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a manufacturer representative (rep) regarding a patient with an implantable neurostimulator (ins).It was reported that there were high impedances on the lead when it was inserted into the ins.The rep stated they inserted the lead into the header of the ins and took impedances and saw > 40,000 ohms on contacts 0, 4, 5, 10, 11, and 15.They removed the lead tails, wiped off the lead and reinserted into the ins and still saw > 40,000.They tried inserting the lead into the wens and saw bad connectivity on (b)(6).They tried this again and saw the same results.The hcp made the decision to replace the lead.The new lead showed normal impedance.No further complications were reported.
 
Additional Manufacturer Narrative
Product id 97715.Analysis results were not available at the time of this report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Analysis of the 977c165 (b)(4)) found no significant anomalies.(b)(4).Below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.The returned device passed all testing in the laboratory and no anomalies were identified.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECIFY SURESCAN
Common Device Name
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key8166160
Report Number2649622-2018-20691
Device Sequence Number16830735
Product Code LGW
UDI-Device Identifier00643169464599
UDI-Public00643169464599
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/25/2021
Device Model Number977C165
Device Catalogue Number977C165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2018
Supplement Date Received by Manufacturer12/12/2018
01/11/2019
Initial Report FDA Received Date12/14/2018
Supplement Report FDA Received Date01/04/2019
02/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
-
-