• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE INC. NUVASIVE PRECEPT SPINAL SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE INC. NUVASIVE PRECEPT SPINAL SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 8800000006
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product has been returned for investigation as product remains in-situ.Radiographs provided confirm a lock screw is not in the fully locked down position.No torque value or device usage confirmed.Unknown if the patient conformed to post-operative activity restriction or suffered a fall.No root cause can be determined as lock screw securing is dependent upon proper rod reduction and torque applied which have not been confirmed.No patient injury reported.Labeling review: ".Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s) loss of fixation." ".All implants should be used only with the appropriately designated instrument (reference surgical technique)." ".Patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." ".Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques." left in-situ.
 
Event or Problem Description
On (b)(6) 2017, patient underwent transforaminal lumbar interbody fusion procedure at l4/5 levels with posterior fixation.Follow up radiographs confirmed a set screw back out at right side of l4 level.A revision procedure is being planned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE PRECEPT SPINAL SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE INC.
7475 lusk blvd
san diego CA 92121
MDR Report Key8166289
Report Number2031966-2018-00173
Device Sequence Number19377752
Product Code NKB
UDI-Device Identifier00887517344076
UDI-Public00887517344076
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K122352
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8800000006
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2018
Initial Report FDA Received Date12/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-