• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER LNCS NEO; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER LNCS NEO; OXIMETER Back to Search Results
Model Number 2329
Medical Device Problem Codes Incorrect Measurement (1383); Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The five returned sensors were evaluated.During evaluation the sensors passed all visual and functional testing.During simulation testing, the sensors were found to provide accurate measurements.No performance issues were identified.The reported issues were unable to be confirmed.
 
Event or Problem Description
The customer reported sensors were found to be faulty on five occasions within the same department and time frame.No consequences or impact to patient were reported.The maude database description of the issue (mw5081615) reads: "five different pts that had either no sao2 reading or wrong sao2 readings with the pulse oximeter probes by masimo, part #2329.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LNCS NEO
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key8166562
Report Number2031172-2018-00577
Device Sequence Number13928955
Product Code DQA
UDI-Device Identifier10843997000922
UDI-Public10843997000922
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K101896
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2021
Device Model Number2329
Device Catalogue Number2329
Device Lot NumberE18JKE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2018
Initial Report FDA Received Date12/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-