ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3262
Device Problem Over-Sensing (1438)
Patient Problem Bradycardia (1751)
Event Date 12/07/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient experienced bradycardia and presented in clinic for follow-up.Upon interrogation, the pacemaker exhibited oversensing of an atrial fibrillation rhythm leading to pacing inhibition.Programming changes were made to the sensing configuration, and the oversensing was resolved.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8167893
MDR Text Key130463669
Report Number2017865-2018-18991
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734509091
UDI-Public05414734509091
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date 12/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Model NumberPM3262
Device Lot NumberA000048704
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age83 YR
-
-