• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523RNAH
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Insulin pump received with broken battery tube thread, reservoir tube lip completely broken off, missing reservoir tube lip o-ring, cracked case reservoir tube window corners and cracked reservoir tube window noted.Unable to perform displacement test due to reservoir lip broken off.The test reservoir does not stay locked in place due to reservoir tube lip broken off.
 
Event or Problem Description
Customer reported via phone call that there was a crack and chipped piece on the battery compartment and the window where the reservoir compartment.Customer's blood glucose level was unknown at the time of the incident.The customer was assisted with troubleshooting.Customer stated that the reservoir was able to lock in place when inserted into insulin pump.The customer was advised to discontinue use of the insulin pump and to revert to the backup plan.The device will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key8168051
Report Number2032227-2018-75437
Device Sequence Number13890066
Product Code OYC
UDI-Device Identifier00643169221529
UDI-Public(01)00643169221529
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523RNAH
Device Catalogue NumberMMT-523RNAH
Device Lot NumberA000291727
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2018
Initial Report FDA Received Date12/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Patient Weight115
-
-