COOK IRELAND LTD ZILVER 635 BILIARY SELF-EXPANDING STENT; FGE CATHETER, BILIARY, DIAGNOSTIC - BILIARY STENT - METAL
|
Back to Search Results |
|
| Model Number |
G36021 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # = k182980.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Device evaluation the zib6-125-4.0-80 device of lot number c1432059 involved in this complaint was returned for evaluation, without the original packaging.With the information provided, a physical examination and document based investigation was conducted.Lab evaluation the device related to this occurrence underwent a laboratory evaluation on the 15 november 2018.The flexor was separated from the handle as a result of using the device off label.Document review prior to distribution all zib6 devices are subject to visual inspection and functional checks to ensure device integrity.A review of the relevant manufacturing records (c1432059) revealed no discrepancies that could have contributed to this complaint.Upon review of complaints, this failure mode has not occurred previously with this lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1432059.It may be noted that the instructions for use (ifu0042-9) states the following: ¿intended use the zilver 518 and 635 biliary stents are intended for palliation of malignant neoplasms in the biliary tree.¿ root cause review a definitive root cause of off label use resulting in handle-flexor separation was identified in the laboratory.The device was used in the superficial femoral artery (sfa).Summary complaint is confirmed as the failure was verified in the laboratory.The flexor was separated from the handle as a result of using the device off label.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Event or Problem Description
|
|
As initially reported to customer relations: for an sfa procedure the zilver 635 biliary self-expanding stent, g36021, was used.Stent deployed, apon removal of delivery system, resistance and delivery system came apart.Follow up customer relation requested how they were able to remover the part of the device that broke.District manager reply per customer: delivery handle came off so they pulled on remaining sheath.(b)(6) 23oct2018.
|
| |
|
Search Alerts/Recalls
|
|
|