| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage(kitchen).Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from all the results obtained.The expected fasting blood glucose test result range is 110-121mg/dl.The customer did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored according to specification in the kitchen.The test strip lot manufacturer's expiration date is 03/19/2020 and open vial date is (b)(6) 2018.The meter memory was reviewed for previous test result history: result 1:164mg/dl date: (b)(6) 2018 time:12:36pm fasting.Result 2:131mg/dl date: (b)(6) 2018 time:12:14pm fasting.Result 3:158mg/dl date: (b)(6) 2018 time:11:26am fasting.Result 4:206mg/dl date: (b)(6) 2018 time:11:13am fasting.
|
| |
|