| Catalog Number |
FDB5.3-35-80 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/25/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
|
| |
|
Event or Problem Description
|
|
The complainant reported that the balloon on the flow-directed balloon catheter would not inflate, so another device of the same kind was used to complete the occlusion procedure.It was further noted that the tip of the balloon had a leak, and the material was pulling away at the distal end of the balloon causing the failure.This product problem did not result in any reported serious injuries to the patient.
|
| |
|