The delivery system was not intact, the proximal end of the pusher has been detached utilizing the secondary break technique.There is no coupler tube or actuator interface attached or returned with the device.The total length of the detached pusher is 179.5cm.The pusher is kinked in several locations along the length at 75.5cm, 129.2cm, 153.9cm, and 179cm from the proximal end.The actuator interface crimp, twr crimp and coupler weld, and ball weld could not be inspected as these portions of the device were not returned.The shield coil is present and in good condition.The release wire is not in the pusher and appears to have been remove during secondary detachment.Dimensions, visual inspection, articulation and liveliness of the release wire and coin could not there fore be verified.The retainer ring was measured to be 0.0046" and is in specification (0.00455"+/- 0.00010").The liner was removed from the distal tip in order to inspect the inner diameter of the lumen stop.Blood was present inside the tip of the pusher and 70% of alcohol was used to swap it clean in order to expose the id of the lumen stop.The lumen stop inside diameter was measured to be 0.0027in (inner fit id is 0.00220" minimum, and for the lumen stop inner fit id is 0.0029" maximum.Both the retainer ring and lumen stop were inner diameters were within dimensional specification and in good visual condition with no signs of damage.No other anomalies were observed.Based on the analysis findings, the root cause of the non-detachment leading to delayed detachment could not be determined.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|