ENDOLOGIX AFX; LIMB STENT GRAFT Back to Search Results
Model Number I20-13/C88F SA
Medical Device Problem Codes Loss of or Failure to Bond (1068); Leak/Splash (1354)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At the completion of the clinical evaluation and based on the information received, there was unsubstantial evidence to support the reported event of type ib endoleak.There was evidence to reasonably support that a non-endologix stent was implanted distal to the left iliac (23mm between stents) per ct scan on (b)(6) 2016 (off-label).Device, user, procedure or anatomy relatedness of this complaint could not be determined.Procedure related harms for this complaint could not be determined.The final patient status was not reported.The manufacturing lot review confirmed all devices met specifications prior to release.This complaint is not capa eligible at this time.These types of events will be monitored and trended as part of the quality system.
 
Event or Problem Description
The patient was initially treated with the afx abdominal aortic aneurysm stent.Approximately one (1) year post initial procedure, a type ib endoleak was detected.The physician elected to treat the patient by implanting an extender ovation ix device and coiling the hypogastric artery on (b)(6) 2016.There have been no additional patient sequelae reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
LIMB STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key8169906
Report Number2031527-2018-00942
Device Sequence Number1326568
Product Code MIH
UDI-Device Identifier00818009011459
UDI-Public(01)00818009011459(17)180821
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/21/2018
Device Model NumberI20-13/C88F SA
Device Lot Number1252611-018
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2016
Initial Report FDA Received Date12/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX-BIFURCATED STENT GRAFT:1251936-001
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-