• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC AUROUS CENTIMETER VESSEL SIZING CATHETER; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC AUROUS CENTIMETER VESSEL SIZING CATHETER; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number N5.0-35-100-P-10S-PIG-CSC-20
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/14/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event or Problem Description
The complainant reported that a hole "blew out" at the proximal end of the aurous centimeter vessel sizing catheter during the final run of the procedure.This part of the catheter was outside of the body when it happened.The procedure was for an aaa (abdominal aortic aneurysm) repair via the groins and abdomen.This was a 6-7 hour procedure, and the catheter had been inserted and removed several times.There was suspicion for a kink at the location of the hole.It was also noted that a power injector was used at 1000 psi.This product problem did not result in any reported serious injuries to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUROUS CENTIMETER VESSEL SIZING CATHETER
Common Device Name
DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8170056
Report Number1820334-2018-03806
Device Sequence Number2142418
Product Code DQO
UDI-Device Identifier00827002119161
UDI-Public(01)00827002119161(17)190509(10)6925529
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/09/2019
Device Catalogue NumberN5.0-35-100-P-10S-PIG-CSC-20
Device Lot Number6925529
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2018
Initial Report FDA Received Date12/17/2018
Patient Sequence Number1
-
-