| Medical Device Problem Code |
Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a button/keypad (tactile changes/unresponsive) issue; the ok button is reportedly not working properly.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the issue can result in inadvertent or incorrect insulin delivery or the inability to use the pump.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 07-feb-2019 with the following findings: during investigation, unable to duplicate complaint about keypad.Peeling keypad from base.There was no other damage to keypad.All keys were responsive.The keypad was removed and there was contamination found under all the key contacts.The pump was opened and there was no evidence of moisture corrosion near the keypad connector.Unrelated to the original complaint, the battery compartment was observed to be cracked.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
| |
|