| Catalog Number |
FS-OMNI-ACRO-35-260 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Continued from concomitant products : cook fusion quattro extraction balloon, fs-qeb-a.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The subassembly device history record for the wire guide subassembly lot number said to be involved was reviewed.Non-conformances were observed that could be related to the reported observation.The device goes through different inspections prior to leaving the facility.These inspections would have removed any nonconforming devices prior to distribution.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide damage.Prior to distribution, all fusion® pre-loaded with acrobat wire guides are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion pre-loaded with acrobat wire guide sphincterotome.The wire coating stripped off from the core.The guide wire coating delaminated [peeled] during the procedure and needed replacing.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Event or Problem Description
|
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion pre-loaded with acrobat wire guide sphincterotome.The wire coating stripped off from the core.The guide wire coating delaminated [peeled] during the procedure and needed replacing.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: cook fusion quattro extraction balloon, fs-qeb-a.Investigation evaluation: our evaluation of the product said to be involved confirmed the report.There was wire guide coating damage near the distal end.The distal 13.5 cm has numerous kinks and some slight discoloration of the silver markings.At 26.7 cm to 28.2 cm from the distal end, was a section of bare core wire.At 28.2 cm, there was a piece of coating that has frayed and was hanging off of the wire guide.Due to the condition of the returned device it cannot be determined if any sections of the coating are missing.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The subassembly device history record for the wire guide subassembly lot number said to be involved was reviewed.Non-conformances were observed that could be related to the reported observation.The device goes through different inspections prior to leaving the facility.These inspections would have removed any nonconforming devices prior to distribution.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide damage.Prior to distribution, all fusion pre-loaded with acrobat wire guide sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|