| Model Number |
42335 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the drainage valve had poor elasticity.The valve was never implanted and there was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was patent.It met the requirements for valve flux, reflux, leak, preimplantation and pressure-flow testing.Therefore, the nature of the complaint could not be replicated by laboratory personnel.There was proteinaceous debris noted within the exterior of the valve.The outlet of the valve was discolored.It is unknown how or when this discoloration occurred.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported the valve did not function well and there was no flow though the valve.It was unknown was caused the issue.
|
| |
|