| Model Number |
9733856R |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No parts have been returned to the manufacturer for analysis.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside of procedure.It was reported that the system cannot load into administrator appliance.The system stopped responding when attempting to load into the screen and a hard restart was required.
|
| |
|
Additional Manufacturer Narrative
|
|
Onsite functional and visual examination was performed by a manufacturer representative.The computer was replaced and the issue was resolved.The system passed a system checkout and was returned to an operational condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|