Model Number SE-0910
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial reporter is synthes sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2018, the patient underwent an akin osteotomy following a lapidus procedure.While the surgeon was proceeding to insert the speed compression implant kit, the staple prematurely came off of the inserter and disrupted cortical bone where predrilled.This premature disengagement of the speed compression implant kit staple was also reported as occurring before.They stated that the current staples seem to have a loading difference because of the pressure needed to release from the inserter.The surgeon had to switch to using cannulated screws instead.Procedure outcome is unknown.Patient status is unknown.This report is for one (1) speed compression implant kit 09x10x10mm.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEED COMPRESSION IMPLANT KIT 09X10X10MM
Common Device Name
BONE STAPLE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8170983
Report Number2939274-2018-55423
Device Sequence Number11199443
Product Code JDR
UDI-Device Identifier00810633020111
UDI-Public(01)00810633020111
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K142292
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSE-0910
Device Catalogue NumberSE-0910
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/26/2018
Initial Report FDA Received Date12/17/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-