The reported event that anterior midshaft plate variax clavicle 10 hole was alleged of 'breakage during surgery' could be confirmed.Based on investigation, the root cause was attributed to be user related.The failure was caused by too much mechanical force which was applied during plate bending.Note; the surgeon completed the case using the broken plate (in conjunction with a hook plate) as this appeared to be adequate for the patient.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|