|
(b)(4).Batch # r9444a.Device analysis: the analysis results found that er320 device was returned inside its package unopened.Upon visual inspection, it was observed that the tyvek from the packaging was damaged.A blue dye test was performed and was confirmed that the sterility was not compromised.Due to the damages found on the packaging, a possible cause for these conditions is due to improper handling during transit or storage; it appears that the package hit a pointy surface and this caused the reported event.The reported complaint was confirmed.All ees product is 100% inspected prior to release.The information you provided is compiled monitored and reviewed on a routine basis for any associated trends.The lot history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the packaging process.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|