ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number B11LT
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Only event year known: 2018.Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.As a lot/batch was not provided, a device history could not be performed.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Was the sterility compromised?.
 
Event or Problem Description
The package uncompleted (have a incision on back).There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Batch # r9444a.Device analysis: the analysis results found that er320 device was returned inside its package unopened.Upon visual inspection, it was observed that the tyvek from the packaging was damaged.A blue dye test was performed and was confirmed that the sterility was not compromised.Due to the damages found on the packaging, a possible cause for these conditions is due to improper handling during transit or storage; it appears that the package hit a pointy surface and this caused the reported event.The reported complaint was confirmed.All ees product is 100% inspected prior to release.The information you provided is compiled monitored and reviewed on a routine basis for any associated trends.The lot history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the packaging process.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key8171096
Report Number3005075853-2018-15173
Device Sequence Number2142437
Product Code GCJ
UDI-Device Identifier10705036001126
UDI-Public10705036001126
Combination Product (Y/N)N
PMA/510(K) Number
K032676
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Catalogue NumberB11LT
Device Lot NumberR40W8U
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2019
Initial Date Received by Manufacturer 12/05/2018
Supplement Date Received by Manufacturer01/04/2019
Initial Report FDA Received Date12/17/2018
Supplement Report FDA Received Date01/18/2019
Patient Sequence Number1
-
-