• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL REDUCTION SPOON T2 FEMUR Ø9 MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL REDUCTION SPOON T2 FEMUR Ø9 MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 18060125
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 10/31/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
The pharmacist at the hospital reported the following event: "i would like to contact you for a declaration of materiovigilance on a guide for nail guide t2 (restérilisable material ref 1806-0125k134075 / 1806-0110k537887).During an intervention on (b)(6) 2018, the screw thread connecting the two parts of the guide broke in the femoral shaft during the attempt to reduce a femoral fracture, resulting in an additional incision on the leg to be able to access the lower end of the instrument in order to be able to remove it.An additional incision had to be made on the leg to gain access to the lower end of the instrument in order to remove it.
 
Additional Manufacturer Narrative
Please note corrections to d4 (catalog number, lot code, gtin).Investigation revealed the subject product to be a concomitant item.The reduction spoon t2 femur did not contribute to the reported event of broken off tip of universal rod t2 femur.
 
Event or Problem Description
The pharmacist at the hospital reported the following event : "i would like to contact you for a declaration of materiovigilance on a guide for nail guide t2 (restérilisable material ref 1806-0125k134075 / 1806-0110k537887).During an intervention on (b)(6) 2018, the screw thread connecting the two parts of the guide broke in the femoral shaft during the attempt to reduce a femoral fracture, resulting in an additional incision on the leg to be able to access the lower end of the instrument in order to be able to remove it.An additional incision had to be made on the leg to gain access to the lower end of the instrument in order to remove it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDUCTION SPOON T2 FEMUR Ø9 MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
MDR Report Key8171155
Report Number0009610622-2018-01514
Device Sequence Number2070338
Product Code LXH
UDI-Device Identifier04546540196019
UDI-Public04546540196019
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 03/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number18060125
Device Lot NumberK134075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2018
Initial Date Received by Manufacturer 11/21/2018
Supplement Date Received by Manufacturer03/06/2019
Initial Report FDA Received Date12/17/2018
Supplement Report FDA Received Date03/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-