| Model Number |
N/A |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The repeat testing was performed using the same sample, without additional centrifugation between initial and repeated measurements.The sample tube was a serum tube with gel separator from greiner bioone.The sample was centrifuged for 10 minutes at 2000xg after a clotting time of 30 minutes.The quality control and adjustment were valid.The performance of other tests was good, so system malfunction is unlikely.The cause for the discordant afp results is unknown.Siemens healthcare diagnostics is investigating.The ifu states in the interpretation of results section: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings".
|
| |
|
Event or Problem Description
|
|
A discordant high advia centaur xp afp result was obtained for a patient sample.The patient sample was repeated and the results were lower.Patient treatment was not prescribed or altered.There was no report of adverse health consequences due to the discordant afp results.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens filed the initial mdr 1219913-2018-00302 on 17-dec-2018.07-jan-2019 additional information: a siemens field service engineer (fse) was sent to the customer site.The fse changed the base pump, tubing of pinch valve and sealing of wash probe.The customer has not observed any further fliers post service.There is no sample available for in house testing.In house kit release was reviewed and no fliers were observed at the time the kit was released.The cause for the discordant afp results is unknown.The instrument is performing within specifications.No further evaluation of the device is required.
|
| |
|
Search Alerts/Recalls
|