Product complaint # (b)(4).The event date was not reported.Procode: krd/hcg.Physical manufacturer name: codman and shurtleff inc., dba depuy synthes products, inc.((b)(4)).Initial reporter - the customer contact information, including name, occupation, phone, fax, and e-mail address, was not reported.The product is not available for evaluation and testing.Additional information will be submitted within 30 days of receipt.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.
|
|
Product complaint # = >(b)(4).[complaint conclusion] as reported by a healthcare professional, the 2.5mm x 4cm ultipaq 10 sr (fsr10025420/ l13863) thermo-mechanical coil failed to detach despite several attempts.The coil was removed from the patient.There was no report of consequence or impact to the patient.Multiple attempts to obtain additional information were unsuccessful.The ultipaq 10 sr, 2.5mmx4cm not returned for evaluation.A review of manufacturing documentation associated with this lot l13863 presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.With the limited information available and without the product available for analysis, the reported customer complaint of failure to detach could not be confirmed.Based on the device history record review, there is no indication that the event is related to the device manufacturing process.Failure to detach is a known potential product failure associated with the use of the device.The ifu contains several cautions relating to this situation, including instructions for troubleshooting the situation should it be encountered during use.Per the ifu: ¿verify that the microcoil delivery system is fully connected, and no faults are indicted on the dcb.If a fault exists, reseat all connections between the dpu, the dcb, and the connecting cable.If a fault still persists, replace the connecting cable.If this does not correct the error, replace the dcb.If the microcoil delivery system still continues to have a fault, retrieve the microcoil, and replace with a new microcoil system.If the system is free of faults, depress the detach button on the blue enpower dcb, the black dcb or the enpower control cable.The detach cycle light next to the button will illuminate and an intermittent tone will sound for the duration of the detachment cycle.If the light and audible tone do not activate, replace the dcb.¿ assignment of root cause for the event remains speculative and inconclusive, based on the minimal information provided and without the return of the complaint device; however, it is possible that procedural and handling factors may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.
|