Catalog Number 1012624-29
Medical Device Problem Codes Difficult to Advance (2920); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual inspection was performed on the returned device.The stent movement on the balloon was confirmed.The resistance was unable to be replicated as it was based on case circumstances.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported revealed no other incidents.The investigation determined that the reported difficulties were due to case circumstances.The reported resistance during advancement and stent movement on the balloon was likely due to anatomical conditions.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat the moderately calcified right common iliac artery.The 8.0mmx29mm omnilink elite stent delivery system (sds) was attempted to be advanced proximally to a previously implanted self-expanding stent.Resistance was felt as the stent entered the target lesion zone, and the stent dislodged from the sds.The stent backed up off the balloon and stayed on the stent/balloon delivery catheter.The catheter was withdrawn with the stent still on the shaft.No snare or additional device was needed for retrieval.A new, same size omnilink elite was used to complete the procedure.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8171419
Report Number2024168-2018-09772
Device Sequence Number16519397
Product Code NIO
UDI-Device Identifier08717648179297
UDI-Public08717648179297
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 12/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Catalogue Number1012624-29
Device Lot Number7112341
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2018
Initial Report FDA Received Date12/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
Patient Weight72
-
-