• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 11.0MM TI HELICAL BLADE 95MM-STERILE; ROD,FIXATION,INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 11.0MM TI HELICAL BLADE 95MM-STERILE; ROD,FIXATION,INTRAMEDULLARY Back to Search Results
Model Number 456.304S
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional product code: hwc.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.A review of the device history records has been requested.The investigation could not be completed; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during a hip surgery on (b)(6) 2018, the helical blade did not insert through the trochanteric fixation nail (tfn) nail congruently.The surgeon removed the nail and implanted another nail with a new helical blade.It was unknown if there was a surgical delay.Procedure and patient outcome is unknown.This complaint involves two (2) devices.This report is 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A device history record (dhr) review was conducted: part number: 456.304s.Lot number: 7510035.Part manufacturing date: 25 october 2013.Manufacturing site: elmira.Part expiration date: 01 september 2022.Nonconformance noted: n/a.A review of the device history record revealed no complaint related anomalies.The device history record shows lot 7510035 of ti helical blades was processed through the normal manufacturing and inspection operations with no rework or nonconformances noted.This lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record(s) determined the raw material lot 7100893 met all specifications with no issues documented that would contribute to this complaint condition.A review of the device history record revealed no complaint related anomalies.This lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.A product investigation was conducted.Device evaluation: investigation flow: device interaction/functional and damage: visual (appearance not as expected).Visual inspection: the returned helical blade (456.304s) was examined and showed surface wear and minor deformation at the distal tip consistent with misalignment.Functional test:the helical blade was able to be inserted into the proximal head element hole of the returned nail (456.415s) and the locking mechanism was able to be engaged as intended.The complaint condition of misalignment was unable to be replicated.Dimensional inspection: inspection of the relevant feature, distal tip, was unable to be completed due to post manufacturing damage.Document/specification review: the relevant drawings, reflecting the current revision and manufactured revisions, were reviewed: during the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.Conclusion: the complaint condition is confirmed as damage was observed which is consistent with a misalignment while attempting to insert a helical blade through the nail.No definitive root cause was able to be determined however misalignments are typically surgical technique related and/or a result of patient anatomy.The system risk document was found to adequately address the complaint allegation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11.0MM TI HELICAL BLADE 95MM-STERILE
Common Device Name
ROD,FIXATION,INTRAMEDULLARY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8171775
Report Number2939274-2018-55430
Device Sequence Number19194158
Product Code HSB
UDI-Device Identifier10886982226709
UDI-Public(01)10886982226709
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K011857
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2022
Device Model Number456.304S
Device Catalogue Number456.304S
Device Lot Number7510035
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2018
Initial Date Received by Manufacturer 11/26/2018
Supplement Date Received by Manufacturer01/14/2019
Initial Report FDA Received Date12/17/2018
Supplement Report FDA Received Date01/15/2019
Patient Sequence Number1
-
-