Model Number DM3500 |
Device Problems
Failure to Interrogate (1332); Pacemaker Found in Back-Up Mode (1440); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/02/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that a patient presented remotely via merlin.Net transmission.Upon review, the insertable cardiac monitor had an alert for device reset.A complete device interrogation was not possible in that condition.The device was attempted with maintenance; however, this resulted in an alert for hardware damage.Device replacement was discussed.
|
|
Event Description
|
New information noted that the device was explanted and replaced on (b)(6) 2019.The patient was stable before and after the procedure.
|
|
Manufacturer Narrative
|
The reported error message and unable to interrogate were confirmed.The device could not communicate when received indicating a reset error.After reloading product code, reset error was reproduced.Further testing indicated memory corruption caused reset error.
|
|